Skip to content
ENQUIRY BASKET · 0
No products yet. Add them with the + on any product, then send the whole list as one enquiry.Browse products

Triglycerides-L

Triglycerides-L, 5 x 25 ml Triglycerides-L, 5 x 25 ml, view 2

Photographed: the 5 x 25 ml pack, stock code CC3-TGS.18N.

2 pack sizes, 125 to 300 ml.

Choose a pack size

Stock code CC3-TGS.18N

  • Made and batch-released in Vadodara
  • ISO 13485:2016 quality system
  • Datasheet published: see every stock code below

Product details.

Intended use: Triglycerides-L (Single Liquid) is a ready-to-use reagent for the quantitative estimation of triglycerides in serum or plasma, by the GPO method. Lipoprotein lipase releases glycerol, which glycerol kinase and glycerol-3-phosphate oxidase convert to hydrogen peroxide; peroxidase then forms a pink quinoneimine dye with 4-aminoantipyrine and chlorophenol, read at 520 nm. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.

The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.

Method
Enzymatic GPO, end point
Sample
Serum
Sample volume
20 µl (0.02 ml)
Wavelength
520 nm (500 to 550 nm, or a green filter)
Linearity
Up to 800 mg/dl
Storage
2 to 8 °C
Full specification19 items
Pack
125 to 300 ml
Pack sizes
2 to choose from
Datasheet
Published
Method
Enzymatic GPO, end point
Sample
Serum: fresh, clear, fasting, no hemolysis; heparin or citrate plasma may be used
Sample volume
20 µl (0.02 ml)
Reagent volume
1.0 ml
Wavelength
520 nm (500 to 550 nm, or a green filter)
Temperature
37 °C
Incubation
10 minutes at 37 °C
Standard
200 mg/dl, included
Reagents
Single liquid reagent, ready to use
Shelf life
18 months at 2 to 8 °C, protected from light
Note
Serum reads about 5 mg/dl higher than plasma
Caution
Contains sodium azide; do not ingest
Storage
2 to 8 °C, until the expiry date on the label
Stock codes
CC3-TGS.18N, CC3-TGS.18NU
Made in
Vadodara, Gujarat, India
Performance1 figure
Linearity
Up to 800 mg/dl; dilute higher samples with 0.9% saline
Procedure at a glance4 steps
  1. Pipette 1.0 ml of reagent into the blank, standard and test tubes; add 0.020 ml of standard and of sample to their tubes.
  2. Mix and incubate at 37 °C for 10 minutes.
  3. Read the test and the standard against the reagent blank at 520 nm.
  4. Triglycerides (mg/dl) = absorbance of test ÷ absorbance of standard × 200; multiply by 0.0114 for mmol/L.

Every stock code.

Triglycerides-L ships in 2 pack sizes, each with its own stock code.

Stock codes and pack sizes for Triglycerides-L
Stock code Pack Photograph Datasheet
CC3-TGS.18N 5 x 25 ml Triglycerides-L, 5 x 25 ml Datasheet PDF, CC3-TGS.18N
CC3-TGS.18NU 5 x 60 ml Triglycerides-L, 5 x 60 ml Datasheet PDF, CC3-TGS.18NU

Questions and answers.

Triglycerides-L: what sample does it need?

The datasheet gives the sample as Serum: fresh, clear, fasting, no hemolysis; heparin or citrate plasma may be used. The sample volume is 20 µl (0.02 ml).

Triglycerides-L: what method does it use?

The method on the datasheet is Enzymatic GPO, end point.

Triglycerides-L: what wavelength is it read at?

According to the datasheet, it is read at 520 nm (500 to 550 nm, or a green filter).

Triglycerides-L: what is the linear range?

The datasheet gives the linear range as Up to 800 mg/dl; dilute higher samples with 0.9% saline.

Triglycerides-L: how should it be stored?

The datasheet gives storage as 2 to 8 °C, until the expiry date on the label.

Triglycerides-L: what pack sizes are available?

It comes in 2 pack sizes, each with its own stock code: 5 x 25 ml (CC3-TGS.18N), 5 x 60 ml (CC3-TGS.18NU).

More clinical chemistry.

All 50 clinical chemistry reagents →

Tell us what
your lab needs.

Enquire now →