Photographed: the 4 x 5 ml pack, stock code CC3-LPS.15M.
SINGLE LIQUID format. supplied as 20 ml.
Stock code CC3-LPS.15M
- Made and batch-released in Vadodara
- ISO 13485:2016 quality system
- Datasheet published: see every stock code below
Product details.
Lipase-L (Single Liquid) is a ready-to-use reagent for the in vitro quantitative determination of lipase (triacylglycerolase) in human serum, by the fixed-time method of Ziegenhorn et al. Lipase, with colipase, hydrolyses triolein to glycerides and fatty acids, and the rate of decrease in turbidity at 340 nm is proportional to lipase activity. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.
The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.
- Method
- Turbidimetric
- Sample
- Non-hemolysed serum
- Wavelength
- 340 nm
- Linearity
- Up to 700 U/L
- Storage
- 2 to 8 °C
- Format
- Single liquid
Full specification19 items
- Category
- Clinical chemistry
- Format
- SINGLE LIQUID
- Pack
- 20 ml
- Datasheet
- Published
- Method
- Turbidimetric, fixed time (decreasing)
- Sample
- Non-hemolysed serum; stable 1 week at 18 to 25 °C, 3 weeks at 2 to 8 °C
- Sample volume
- 0.04 ml, or 0.02 ml in the half-volume procedure
- Reagent volume
- 1.0 ml, or 0.5 ml
- Wavelength
- 340 nm
- Temperature
- 37 °C
- Readings
- At 240 seconds and 540 seconds
- Calibration
- Lipase calibrator included; reconstitute with 1 ml of diluent
- Reagents
- Single liquid reagent, ready to use
- Calibrator stability
- 7 days at 2 to 8 °C; 4 weeks at −20 °C
- Note
- Keep cholesterol and triglyceride reagents away: they contain lipase and esterase
- Caution
- Avoid ingestion and contact with skin, mouth and eyes
- Storage
- 2 to 8 °C, until the expiry date on the label
- Stock codes
- CC3-LPS.15M
- Made in
- Vadodara, Gujarat, India
Performance2 figures
- Linearity
- Up to 700 U/L
- Interference
- None from hemoglobin up to 124 µmol/L, bilirubin up to 342 µmol/L
Procedure at a glance5 steps
- Reconstitute the lipase calibrator with 1 ml of diluent.
- Pipette 1.0 ml of reagent into the calibrator and test tubes; add 0.04 ml of calibrator and of sample to their tubes.
- Mix and read the absorbance at 340 nm at exactly 240 seconds and again at 540 seconds.
- Work out ΔA = A1 − A2.
- Lipase (U/L) = ΔA of sample ÷ ΔA of calibrator × calibrator value.
Every stock code.
Lipase-L (Single Liquid) ships in one pack size.
| Stock code | Pack | Photograph | Datasheet |
|---|---|---|---|
| CC3-LPS.15M | 4 x 5 ml | ![]() |
Datasheet |
More clinical chemistry.
All 51 in clinical chemistry →