LDL-Direct, 20 ml / 50 Test with calibrator LDL-Direct, 20 ml / 50 Test with calibrator, view 2

Photographed: the 20 ml / 50 Test with calibrator pack, stock code CC3-LDL.022.

4 pack sizes, 50 to 200 tests · 56 ml.

Choose a pack size

Stock code CC3-LDL.022

  • Supplied by Reckon Diagnostics, Vadodara
  • Datasheet published: see every stock code below

Product details.

Intended use: LDL-Direct (Selective Detergent) is a two-reagent assay for the quantitative determination of LDL cholesterol, with no pre-treatment or centrifugation. The first step eliminates cholesterol from non-LDL lipoproteins; the second measures LDL cholesterol enzymatically, the colour formed being proportional to its concentration. Supplied by Reckon Diagnostics, Vadodara.

The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.

Method
Selective detergent
Sample
Serum, tested without delay after separation
Sample volume
5 µl
Wavelength
578 nm
Linearity
Up to 610 mg/dl
Storage
2 to 8 °C
Full specification18 items
Pack
50 to 200 tests · 56 ml
Pack sizes
4 to choose from
Datasheet
Published
Method
Selective detergent, fixed time; no pre-treatment
Sample
Serum, tested without delay after separation; stable 7 days at 2 to 8 °C; avoid repeated freezing
Sample volume
5 µl
Reagent volume
375 µl of R1 + 125 µl of R2
Wavelength
578 nm
Temperature
37 °C
Incubation
5 minutes with R1, then 5 minutes after R2
Calibration
HDL/LDL calibrator included (lyophilised human serum)
Reagents
R1 and R2, ready to use
Opened vial
R1 and R2: 4 weeks at 2 to 8 °C
Discard if
Particles or turbidity appear
Storage
2 to 8 °C, tightly closed, until the expiry date
Stock codes
CC3-LDL.022, CC3-LDL.22M, CC3-LDL.22MU, SYT-LDL.022
Supplied by
Reckon Diagnostics, Vadodara
Performance3 figures
Linearity
Up to 610 mg/dl; dilute higher samples with isotonic saline
Precision
Intra-assay CV 0.5 to 0.8%; inter-assay CV 1.1 to 1.5%
Interference
None up to ascorbic acid 50 mg/dl, hemoglobin 500 mg/dl, bilirubin 30 mg/dl
Procedure at a glance4 steps
  1. Mix 5 µl of calibrator or sample with 375 µl of R1 and incubate for 5 minutes at 37 °C.
  2. Add 125 µl of R2, mix, and read absorbance A1 after 5 seconds.
  3. Incubate for 5 minutes at 37 °C and read absorbance A2 at 578 nm.
  4. LDL cholesterol (mg/dl) = (A2 − A1) of sample ÷ (A2 − A1) of calibrator × calibrator value.

Every stock code.

LDL-Direct ships in 4 pack sizes, each with its own stock code.

Stock codes and pack sizes for LDL-Direct
Stock code Pack Photograph Datasheet
CC3-LDL.022 20 ml / 50 Test with calibrator LDL-Direct, 20 ml / 50 Test with calibrator Datasheet PDF, CC3-LDL.022
CC3-LDL.22M 40 ml / 100 Test with calibrator LDL-Direct, 40 ml / 100 Test with calibrator Datasheet PDF, CC3-LDL.22M
CC3-LDL.22MU 80 ml / 200 Test with calibrator LDL-Direct, 80 ml / 200 Test with calibrator Datasheet PDF, CC3-LDL.22MU
SYT-LDL.022 R1 4 x 10 ml + R2 4 x 4 ml (system pack for ReDi 120/240) – –

Questions and answers.

LDL-Direct: what sample does it need?

The datasheet gives the sample as Serum, tested without delay after separation; stable 7 days at 2 to 8 °C; avoid repeated freezing. The sample volume is 5 µl.

LDL-Direct: what method does it use?

The method on the datasheet is Selective detergent, fixed time; no pre-treatment.

LDL-Direct: what wavelength is it read at?

According to the datasheet, it is read at 578 nm.

LDL-Direct: what is the linear range?

The datasheet gives the linear range as Up to 610 mg/dl; dilute higher samples with isotonic saline.

LDL-Direct: how should it be stored?

The datasheet gives storage as 2 to 8 °C, tightly closed, until the expiry date.

LDL-Direct: what pack sizes are available?

It comes in 4 pack sizes, each with its own stock code: 20 ml / 50 Test with calibrator (CC3-LDL.022), 40 ml / 100 Test with calibrator (CC3-LDL.22M), 80 ml / 200 Test with calibrator (CC3-LDL.22MU), R1 4 x 10 ml + R2 4 x 4 ml (system pack for ReDi 120/240) (SYT-LDL.022).