Photographed: the 20 ml / 50 Test with calibrator pack, stock code CC3-LDL.022.
4 pack sizes, 50 to 200 tests · 56 ml.
Stock code CC3-LDL.022
- Supplied by Reckon Diagnostics, Vadodara
- Datasheet published: see every stock code below
Product details.
Intended use: LDL-Direct (Selective Detergent) is a two-reagent assay for the quantitative determination of LDL cholesterol, with no pre-treatment or centrifugation. The first step eliminates cholesterol from non-LDL lipoproteins; the second measures LDL cholesterol enzymatically, the colour formed being proportional to its concentration. Supplied by Reckon Diagnostics, Vadodara.
The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.
- Method
- Selective detergent
- Sample
- Serum, tested without delay after separation
- Sample volume
- 5 µl
- Wavelength
- 578 nm
- Linearity
- Up to 610 mg/dl
- Storage
- 2 to 8 °C
Full specification18 items
- Category
- Clinical chemistry
- Pack
- 50 to 200 tests · 56 ml
- Pack sizes
- 4 to choose from
- Datasheet
- Published
- Method
- Selective detergent, fixed time; no pre-treatment
- Sample
- Serum, tested without delay after separation; stable 7 days at 2 to 8 °C; avoid repeated freezing
- Sample volume
- 5 µl
- Reagent volume
- 375 µl of R1 + 125 µl of R2
- Wavelength
- 578 nm
- Temperature
- 37 °C
- Incubation
- 5 minutes with R1, then 5 minutes after R2
- Calibration
- HDL/LDL calibrator included (lyophilised human serum)
- Reagents
- R1 and R2, ready to use
- Opened vial
- R1 and R2: 4 weeks at 2 to 8 °C
- Discard if
- Particles or turbidity appear
- Storage
- 2 to 8 °C, tightly closed, until the expiry date
- Stock codes
- CC3-LDL.022, CC3-LDL.22M, CC3-LDL.22MU, SYT-LDL.022
- Supplied by
- Reckon Diagnostics, Vadodara
Performance3 figures
- Linearity
- Up to 610 mg/dl; dilute higher samples with isotonic saline
- Precision
- Intra-assay CV 0.5 to 0.8%; inter-assay CV 1.1 to 1.5%
- Interference
- None up to ascorbic acid 50 mg/dl, hemoglobin 500 mg/dl, bilirubin 30 mg/dl
Procedure at a glance4 steps
- Mix 5 µl of calibrator or sample with 375 µl of R1 and incubate for 5 minutes at 37 °C.
- Add 125 µl of R2, mix, and read absorbance A1 after 5 seconds.
- Incubate for 5 minutes at 37 °C and read absorbance A2 at 578 nm.
- LDL cholesterol (mg/dl) = (A2 − A1) of sample ÷ (A2 − A1) of calibrator × calibrator value.
Every stock code.
LDL-Direct ships in 4 pack sizes, each with its own stock code.
| Stock code | Pack | Photograph | Datasheet |
|---|---|---|---|
| CC3-LDL.022 | 20 ml / 50 Test with calibrator | ![]() |
Datasheet PDF, CC3-LDL.022 |
| CC3-LDL.22M | 40 ml / 100 Test with calibrator | ![]() |
Datasheet PDF, CC3-LDL.22M |
| CC3-LDL.22MU | 80 ml / 200 Test with calibrator | ![]() |
Datasheet PDF, CC3-LDL.22MU |
| SYT-LDL.022 | R1 4 x 10 ml + R2 4 x 4 ml (system pack for ReDi 120/240) | – | – |
Questions and answers.
LDL-Direct: what sample does it need?
The datasheet gives the sample as Serum, tested without delay after separation; stable 7 days at 2 to 8 °C; avoid repeated freezing. The sample volume is 5 µl.
LDL-Direct: what method does it use?
The method on the datasheet is Selective detergent, fixed time; no pre-treatment.
LDL-Direct: what wavelength is it read at?
According to the datasheet, it is read at 578 nm.
LDL-Direct: what is the linear range?
The datasheet gives the linear range as Up to 610 mg/dl; dilute higher samples with isotonic saline.
LDL-Direct: how should it be stored?
The datasheet gives storage as 2 to 8 °C, tightly closed, until the expiry date.
LDL-Direct: what pack sizes are available?
It comes in 4 pack sizes, each with its own stock code: 20 ml / 50 Test with calibrator (CC3-LDL.022), 40 ml / 100 Test with calibrator (CC3-LDL.22M), 80 ml / 200 Test with calibrator (CC3-LDL.22MU), R1 4 x 10 ml + R2 4 x 4 ml (system pack for ReDi 120/240) (SYT-LDL.022).