Photographed: the 20 ml / 50 Test with calibrator pack, stock code CC3-LDL.022.
3 pack sizes, 50 to 200 tests.
Stock code CC3-LDL.022
- Made and batch-released in Vadodara
- ISO 13485:2016 quality system
- Datasheet published: see every stock code below
Product details.
LDL-Direct (Selective Detergent) is a two-reagent assay for the quantitative determination of LDL cholesterol, with no pre-treatment or centrifugation. The first step eliminates cholesterol from non-LDL lipoproteins; the second measures LDL cholesterol enzymatically, the colour formed being proportional to its concentration. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.
The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.
- Method
- Selective detergent
- Sample
- Serum, tested without delay after separation
- Sample volume
- 5 µl
- Wavelength
- 578 nm
- Linearity
- Up to 610 mg/dl
- Storage
- 2 to 8 °C
Full specification18 items
- Category
- Clinical chemistry
- Pack
- 50 to 200 tests
- Pack sizes
- 3 to choose from
- Datasheet
- Published
- Method
- Selective detergent, fixed time; no pre-treatment
- Sample
- Serum, tested without delay after separation; stable 7 days at 2 to 8 °C; avoid repeated freezing
- Sample volume
- 5 µl
- Reagent volume
- 375 µl of R1 + 125 µl of R2
- Wavelength
- 578 nm
- Temperature
- 37 °C
- Incubation
- 5 minutes with R1, then 5 minutes after R2
- Calibration
- HDL/LDL calibrator included (lyophilised human serum)
- Reagents
- R1 and R2, ready to use
- Opened vial
- R1 and R2: 4 weeks at 2 to 8 °C
- Discard if
- Particles or turbidity appear
- Storage
- 2 to 8 °C, tightly closed, until the expiry date
- Stock codes
- CC3-LDL.022, CC3-LDL.22M, CC3-LDL.22MU
- Made in
- Vadodara, Gujarat, India
Performance3 figures
- Linearity
- Up to 610 mg/dl; dilute higher samples with isotonic saline
- Precision
- Intra-assay CV 0.5 to 0.8%; inter-assay CV 1.1 to 1.5%
- Interference
- None up to ascorbic acid 50 mg/dl, hemoglobin 500 mg/dl, bilirubin 30 mg/dl
Procedure at a glance4 steps
- Mix 5 µl of calibrator or sample with 375 µl of R1 and incubate for 5 minutes at 37 °C.
- Add 125 µl of R2, mix, and read absorbance A1 after 5 seconds.
- Incubate for 5 minutes at 37 °C and read absorbance A2 at 578 nm.
- LDL cholesterol (mg/dl) = (A2 − A1) of sample ÷ (A2 − A1) of calibrator × calibrator value.
Every stock code.
LDL-Direct ships in 3 pack sizes, each with its own stock code.
| Stock code | Pack | Photograph | Datasheet |
|---|---|---|---|
| CC3-LDL.022 | 20 ml / 50 Test with calibrator | ![]() |
Datasheet |
| CC3-LDL.22M | 40 ml / 100 Test with calibrator | ![]() |
Datasheet |
| CC3-LDL.22MU | 80 ml / 200 Test with calibrator | ![]() |
Datasheet |
More clinical chemistry.