Photographed: the 2 x 25 ml pack, stock code CC3-AST.16M.
R1-R2 format. 2 pack sizes, 50 to 100 ml.
Stock code CC3-AST.16M
- Made and batch-released in Vadodara
- ISO 13485:2016 quality system
- Datasheet published: see every stock code below
Product details.
SGOT-L (R1-R2) is a liquid reagent kit for the quantitative estimation of glutamate oxaloacetate transaminase (AST) activity in serum or plasma, by the kinetic procedure recommended by the IFCC. AST forms oxaloacetate from L-aspartate and α-ketoglutarate; malate dehydrogenase reduces it while oxidising NADH, and the fall in absorbance at 340 nm is proportional to AST activity. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.
The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.
- Method
- Kinetic, IFCC (decreasing absorbance)
- Sample
- Serum
- Sample volume
- 50 µl (0.05 ml)
- Wavelength
- 340 nm
- Linearity
- Up to 500 IU/L
- Storage
- 2 to 8 °C
Full specification19 items
- Category
- Clinical chemistry
- Format
- R1-R2
- Pack
- 50 to 100 ml
- Pack sizes
- 2 to choose from
- Datasheet
- Published
- Method
- Kinetic, IFCC (decreasing absorbance)
- Sample
- Serum: fresh, clear, fasting, no hemolysis; heparin or EDTA plasma may be used
- Sample volume
- 50 µl (0.05 ml)
- Reagent volume
- 1.0 ml of working reagent, or 0.8 ml of R1 + 0.2 ml of R2
- Wavelength
- 340 nm
- Temperature
- 37 °C
- Readings
- 4, at 30-second intervals after a 60-second delay
- Factor
- 3376
- Reagents
- R1 (enzyme) and R2 (substrate), liquid; working reagent is 4 parts R1 to 1 part R2
- Working reagent
- 30 days at 2 to 8 °C
- Caution
- Contains sodium azide; do not ingest
- Storage
- 2 to 8 °C, until the expiry date on the label
- Stock codes
- CC3-AST.16M, CC3-AST.16MU
- Made in
- Vadodara, Gujarat, India
Performance1 figure
- Linearity
- Up to 500 IU/L; dilute higher samples with 0.9% saline
Procedure at a glance5 steps
- Pour 1 bottle of R2 into the R1 bottle and wait 2 minutes, or mix 4 parts of R1 with 1 part of R2 as needed.
- Add 0.05 ml of sample to 1.0 ml of working reagent (or to 0.8 ml of R1 and 0.2 ml of R2) and mix.
- After 60 seconds at 37 °C, read the absorbance at 340 nm every 30 seconds for 2 minutes.
- Work out the mean change in absorbance per minute (ΔA/min).
- SGOT (IU/L) = ΔA/min × 3376.
Every stock code.
SGOT-L (R1-R2) ships in 2 pack sizes, each with its own stock code.
| Stock code | Pack | Photograph | Datasheet |
|---|---|---|---|
| CC3-AST.16M | 2 x 25 ml | ![]() |
Datasheet |
| CC3-AST.16MU | 4 x 25 ml | ![]() |
Datasheet |
More clinical chemistry.