Phosphorus-L (Single Liquid)

Phosphorus-L (Single Liquid), 20 Test Phosphorus-L (Single Liquid), 20 Test, view 2

Photographed: the 20 Test pack, stock code CC1-PIC.044.

Single liquid format. Supplied as 20 tests.

Pack size

Stock code CC1-PIC.044

  • Made and batch-released in Vadodara
  • ISO 13485:2016 quality system
  • Datasheet published: see every stock code below

Product details.

Intended use: Phosphorus-L (Single Liquid) is a ready-to-use reagent kit for the quantitative estimation of phosphorus in serum. Phosphorus reacts with ammonium molybdate in sulphuric acid to form an unreduced phosphomolybdate complex, measured at 340 nm. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.

The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.

Method
Ammonium molybdate (unreduced)
Sample
Fresh, clear serum
Sample volume
10 µl (0.01 ml)
Wavelength
340 nm
Linearity
Up to 15 mg/dl
Storage
2 to 8 °C
Full specification17 items
Format
SINGLE LIQUID
Pack
20 tests
Datasheet
Published
Method
Ammonium molybdate (unreduced), end point
Sample
Fresh, clear serum; plasma not recommended
Sample volume
10 µl (0.01 ml)
Reagent volume
1.0 ml
Wavelength
340 nm
Temperature
30 °C
Incubation
5 minutes at room temperature
Standard
5 mg/dl, included
Reagents
Single liquid reagent, ready to use
Note
Decontaminate glassware in 1 N HCl and rinse with distilled water
Storage
2 to 8 °C, until the expiry date on the label
Stock codes
CC1-PIC.044
Made in
Vadodara, Gujarat, India
Performance1 figure
Linearity
Up to 15 mg/dl; dilute higher samples with distilled water
Procedure at a glance4 steps
  1. Pipette 1.0 ml of reagent into the blank, standard and test tubes; add 0.01 ml of standard and of sample to their tubes.
  2. Mix and incubate for 5 minutes at room temperature.
  3. Read the standard and the test against the blank at 340 nm.
  4. Phosphorus (mg/dl) = absorbance of test ÷ absorbance of standard × 5.

Every stock code.

Phosphorus-L (Single Liquid) ships in one pack size.

Stock codes and pack sizes for Phosphorus-L (Single Liquid)
Stock code Pack Photograph Datasheet
CC1-PIC.044 20 Test Phosphorus-L (Single Liquid), 20 Test Datasheet PDF, CC1-PIC.044

Questions and answers.

Phosphorus-L (Single Liquid): what sample does it need?

The datasheet gives the sample as Fresh, clear serum; plasma not recommended. The sample volume is 10 µl (0.01 ml).

Phosphorus-L (Single Liquid): what method does it use?

The method on the datasheet is Ammonium molybdate (unreduced), end point.

Phosphorus-L (Single Liquid): what wavelength is it read at?

According to the datasheet, it is read at 340 nm.

Phosphorus-L (Single Liquid): what is the linear range?

The datasheet gives the linear range as Up to 15 mg/dl; dilute higher samples with distilled water.

Phosphorus-L (Single Liquid): how should it be stored?

The datasheet gives storage as 2 to 8 °C, until the expiry date on the label.

Phosphorus-L (Single Liquid): what pack sizes are available?

It comes in one pack size: 20 Test (CC1-PIC.044).