Photographed: the 20 Test pack, stock code CC1-PIC.044.
Single liquid format. Supplied as 20 tests.
Stock code CC1-PIC.044
- Made and batch-released in Vadodara
- ISO 13485:2016 quality system
- Datasheet published: see every stock code below
Product details.
Intended use: Phosphorus-L (Single Liquid) is a ready-to-use reagent kit for the quantitative estimation of phosphorus in serum. Phosphorus reacts with ammonium molybdate in sulphuric acid to form an unreduced phosphomolybdate complex, measured at 340 nm. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.
The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.
- Method
- Ammonium molybdate (unreduced)
- Sample
- Fresh, clear serum
- Sample volume
- 10 µl (0.01 ml)
- Wavelength
- 340 nm
- Linearity
- Up to 15 mg/dl
- Storage
- 2 to 8 °C
Full specification17 items
- Category
- Clinical chemistry
- Format
- SINGLE LIQUID
- Pack
- 20 tests
- Datasheet
- Published
- Method
- Ammonium molybdate (unreduced), end point
- Sample
- Fresh, clear serum; plasma not recommended
- Sample volume
- 10 µl (0.01 ml)
- Reagent volume
- 1.0 ml
- Wavelength
- 340 nm
- Temperature
- 30 °C
- Incubation
- 5 minutes at room temperature
- Standard
- 5 mg/dl, included
- Reagents
- Single liquid reagent, ready to use
- Note
- Decontaminate glassware in 1 N HCl and rinse with distilled water
- Storage
- 2 to 8 °C, until the expiry date on the label
- Stock codes
- CC1-PIC.044
- Made in
- Vadodara, Gujarat, India
Performance1 figure
- Linearity
- Up to 15 mg/dl; dilute higher samples with distilled water
Procedure at a glance4 steps
- Pipette 1.0 ml of reagent into the blank, standard and test tubes; add 0.01 ml of standard and of sample to their tubes.
- Mix and incubate for 5 minutes at room temperature.
- Read the standard and the test against the blank at 340 nm.
- Phosphorus (mg/dl) = absorbance of test ÷ absorbance of standard × 5.
Every stock code.
Phosphorus-L (Single Liquid) ships in one pack size.
| Stock code | Pack | Photograph | Datasheet |
|---|---|---|---|
| CC1-PIC.044 | 20 Test | ![]() |
Datasheet PDF, CC1-PIC.044 |
Questions and answers.
Phosphorus-L (Single Liquid): what sample does it need?
The datasheet gives the sample as Fresh, clear serum; plasma not recommended. The sample volume is 10 µl (0.01 ml).
Phosphorus-L (Single Liquid): what method does it use?
The method on the datasheet is Ammonium molybdate (unreduced), end point.
Phosphorus-L (Single Liquid): what wavelength is it read at?
According to the datasheet, it is read at 340 nm.
Phosphorus-L (Single Liquid): what is the linear range?
The datasheet gives the linear range as Up to 15 mg/dl; dilute higher samples with distilled water.
Phosphorus-L (Single Liquid): how should it be stored?
The datasheet gives storage as 2 to 8 °C, until the expiry date on the label.
Phosphorus-L (Single Liquid): what pack sizes are available?
It comes in one pack size: 20 Test (CC1-PIC.044).
More clinical chemistry.
All 50 clinical chemistry reagents →