Drawing, not a photograph. This pack has not been photographed yet; the stock codes and pack sizes below are the specification.
Supplied as 20 tests.
Stock code CC1-IRN.055
- Made and batch-released in Vadodara
- ISO 13485:2016 quality system
- Datasheet published: see every stock code below
Product details.
Iron (Ferrozine) is a reagent kit for the quantitative estimation of iron in serum. In an acidic medium, transferrin-bound iron is released as ferric ions and reduced by hydroxylamine to ferrous ions, which form a purple complex with ferrozine measured at 560 nm. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.
The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.
- Method
- Ferrozine, end point (increase)
- Sample
- Serum
- Sample volume
- 200 µl (0.200 ml)
- Wavelength
- 560 nm
- Linearity
- Up to 89 µmol/L (500 µg/dl)
- Storage
- 2 to 8 °C
Full specification18 items
- Category
- Clinical chemistry
- Pack
- 20 tests
- Datasheet
- Published
- Method
- Ferrozine, end point (increase)
- Sample
- Serum: fresh, clear, no hemolysis; not plasma; stable 1 day at room temperature or 1 week at 2 to 8 °C
- Sample volume
- 200 µl (0.200 ml)
- Reagent volume
- 1.0 ml of working reagent, then 0.05 ml of colour reagent
- Wavelength
- 560 nm; bichromatic 560 and 630 nm
- Temperature
- 37 °C
- Incubation
- 10 minutes at 37 °C
- Limitation
- Hemolysis gives falsely high results; iron medication affects serum levels
- Standard
- 80 µmol/L (446 µg/dl), included
- Reagents
- Buffer, colour and enhancer reagents; 1 vial of enhancer in 1 bottle of buffer, used after 15 minutes
- Working reagent
- 60 days at 2 to 8 °C
- Note
- Glassware must be iron-free: soak in 0.1 N HNO3 or 1 N HCl and rinse with deionised water
- Storage
- 2 to 8 °C, protected from light
- Stock codes
- CC1-IRN.055
- Made in
- Vadodara, Gujarat, India
Performance1 figure
- Linearity
- Up to 89 µmol/L (500 µg/dl); dilute higher samples with 0.9% saline
Procedure at a glance5 steps
- Dissolve 1 vial of 3 Iron (enhancer) in 1 bottle of 1 Iron (buffer) and wait 15 minutes.
- Pipette 1.0 ml of working reagent into the blank, standard, sample-blank and test tubes; add 0.2 ml of distilled water, standard, and sample (to both sample tubes).
- Add 0.05 ml of 2 Iron (colour reagent) to the blank, standard and test tubes, not to the sample blank.
- Stand for 10 minutes at 37 °C and read against the reagent blank at 560 nm.
- Iron (µmol/L) = (test − sample blank) ÷ standard × 80; multiply by 5.585 for µg/dl.
Every stock code.
Iron ships in one pack size.
| Stock code | Pack | Photograph | Datasheet |
|---|---|---|---|
| CC1-IRN.055 | 20 Test | — | Datasheet |
More clinical chemistry.
All 51 in clinical chemistry →