Skip to content
ENQUIRY BASKET · 0
Empty. Add products with the + on any card, then send the whole list as one enquiry.
CLINICAL CHEMISTRY · HDL-DIRECT
HDL-Direct, 20 ml / 50 Test with calibrator

Photographed: the 20 ml / 50 Test with calibrator pack, stock code CC3-HDL.13M.

3 pack sizes, 50 to 200 tests.

Choose a pack size

Stock code CC3-HDL.13M

ASK ABOUT THIS PRODUCT →
  • Made and batch-released in Vadodara
  • ISO 13485:2016 quality system
  • Datasheet published: see every stock code below

Product details.

HDL-Direct (Selective Detergent) is a two-reagent assay for the quantitative determination of HDL cholesterol, with no precipitation step. The first reagent consumes non-HDL cholesterol in a colourless reaction; the second solubilises HDL selectively and develops colour in proportion to HDL cholesterol. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.

The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.

Method
Selective detergent
Sample
Serum, EDTA or heparin plasma
Sample volume
10 µl semi-automated
Linearity
Up to 150 mg/dl (4.0 mmol/L)
Storage
2 to 8 °C
Pack
50 to 200 tests
Full specification17 items
Pack
50 to 200 tests
Pack sizes
3 to choose from
Datasheet
Published
Method
Selective detergent, fixed time; no precipitation
Sample
Serum, EDTA or heparin plasma; fasting not needed; freeze at −80 °C if kept beyond 5 days
Sample volume
10 µl semi-automated; 3 µl fully automated
Reagent volume
750 µl of R1 + 250 µl of R2 semi-automated; 225 µl + 75 µl fully automated
Wavelength
578 nm (700 nm secondary on fully automated analyzers)
Temperature
37 °C
Incubation
5 minutes with R1, then 5 minutes after R2
Calibration
HDL/LDL calibrator included (lyophilised human serum)
Reagents
R1 and R2, ready to use
Opened vial
R1 and R2: 30 days at 2 to 8 °C
Storage
2 to 8 °C, tightly closed, until the expiry date
Stock codes
CC3-HDL.13M, CC3-HDL.13MU, CC3-HDL.13MV
Made in
Vadodara, Gujarat, India
Performance3 figures
Linearity
Up to 150 mg/dl (4.0 mmol/L); dilute higher samples with isotonic saline
Precision
Intra-assay CV 0.65 to 0.78%; inter-assay CV 1.11 to 1.33%
Interference
None up to ascorbic acid 100 mg/dl, bilirubin 40 mg/dl, hemoglobin 500 mg/dl
Procedure at a glance4 steps
  1. Mix 10 µl of calibrator or sample with 750 µl of R1 and incubate for 5 minutes at 37 °C.
  2. Add 250 µl of R2, mix, and read absorbance A1 after 5 seconds.
  3. Incubate for 5 minutes at 37 °C and read absorbance A2 at 578 nm.
  4. HDL cholesterol (mg/dl) = (A2 − A1) of sample ÷ (A2 − A1) of calibrator × calibrator value.

Every stock code.

HDL-Direct ships in 3 pack sizes, each with its own stock code.

Stock codes and pack sizes for HDL-Direct
Stock code Pack Photograph Datasheet
CC3-HDL.13M 20 ml / 50 Test with calibrator HDL-Direct, 20 ml / 50 Test with calibrator Datasheet
CC3-HDL.13MU 40 ml / 100 Test with calibrator HDL-Direct, 40 ml / 100 Test with calibrator Datasheet
CC3-HDL.13MV 80 ml / 200 Test with calibrator HDL-Direct, 80 ml / 200 Test with calibrator Datasheet

More clinical chemistry.

All 51 in clinical chemistry →

Tell us what
your lab needs.

ENQUIRE NOW →