Photographed: the 2 x 25 ml pack, stock code CC2-HDC.013.
Supplied as 50 ml.
Stock code CC2-HDC.013
- Made and batch-released in Vadodara
- ISO 13485:2016 quality system
- Datasheet published: see every stock code below
Product details.
HDL-Cholesterol (PTA, Mg++/CHOD-PAP) is a reagent kit for the quantitative estimation of high density lipoprotein (HDL) cholesterol in serum or plasma. Phosphotungstic acid and magnesium chloride precipitate the other lipoprotein fractions, and the cholesterol left in the clear supernatant is measured enzymatically at 505 nm. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.
The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.
- Method
- PTA/Mg++ precipitation
- Sample
- Serum
- Sample volume
- 50 µl (0.05 ml) of supernatant
- Wavelength
- 505 nm (500 to 540 nm, or a green filter)
- Linearity
- Up to 250 mg/dl
- Storage
- 2 to 8 °C
Full specification18 items
- Category
- Clinical chemistry
- Pack
- 50 ml
- Datasheet
- Published
- Method
- PTA/Mg++ precipitation, then CHOD-PAP end point
- Sample
- Serum: fresh, clear, fasting, no hemolysis; heparin plasma may also be used
- Precipitation
- 0.2 ml of sample with 0.2 ml of precipitating reagent; centrifuge 10 minutes at 3500 to 4000 rpm
- Sample volume
- 50 µl (0.05 ml) of supernatant
- Reagent volume
- 1.0 ml of working reagent
- Wavelength
- 505 nm (500 to 540 nm, or a green filter)
- Temperature
- 37 °C
- Incubation
- 10 minutes at 37 °C
- Standard
- 50 mg/dl, included; not put through the precipitation step
- Reagents
- Enzyme-chromogen, diluent, and a ready-to-use precipitating reagent
- Working reagent
- 180 days at 2 to 8 °C
- Caution
- Contains sodium azide; do not ingest
- Storage
- 2 to 8 °C, until the expiry date on the label
- Stock codes
- CC2-HDC.013
- Made in
- Vadodara, Gujarat, India
Performance1 figure
- Linearity
- Up to 250 mg/dl; dilute higher samples with 0.9% saline
Procedure at a glance5 steps
- Mix 0.2 ml of sample with 0.2 ml of precipitating reagent, centrifuge at 3500 to 4000 rpm for 10 minutes, and separate the clear supernatant straight away.
- Dissolve 1 vial of enzyme-chromogen in the diluent and wait 10 minutes.
- Pipette 1.0 ml of working reagent into the blank, standard and test tubes; add 0.05 ml of standard and of supernatant to their tubes.
- Mix, incubate for 10 minutes at 37 °C, and read against the reagent blank at 505 nm.
- HDL cholesterol (mg/dl) = absorbance of test ÷ absorbance of standard × 50 × 2.