Cholesterol-L (Single Liquid)

Cholesterol-L (Single Liquid), 2 x 60 ml Cholesterol-L (Single Liquid), 2 x 60 ml, view 2

Photographed: the 2 x 60 ml pack, stock code CC3-CLE.05M.

Single liquid format. 2 pack sizes, 120 to 300 ml.

Choose a pack size

Stock code CC3-CLE.05M

  • Made and batch-released in Vadodara
  • ISO 13485:2016 quality system
  • Datasheet published: see every stock code below

Product details.

Intended use: Cholesterol-L (Single Liquid) is a ready-to-use liquid reagent for the quantitative estimation of cholesterol in serum or plasma. Cholesterol esterase and cholesterol oxidase release hydrogen peroxide, which with peroxidase, 4-aminoantipyrine and a phenolic compound forms a quinoneimine dye read at 505 nm in proportion to total cholesterol. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.

The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.

Method
Enzymatic CHOD-PAP, end point
Sample
Serum
Sample volume
10 µl
Wavelength
505 nm (505 to 530 nm, or a green filter)
Linearity
Up to 1000 mg/dl
Storage
2 to 8 °C
Full specification18 items
Format
SINGLE LIQUID
Pack
120 to 300 ml
Pack sizes
2 to choose from
Datasheet
Published
Method
Enzymatic CHOD-PAP, end point
Sample
Serum: fresh, clear, fasting, no hemolysis; heparin or EDTA plasma may also be used
Sample volume
10 µl
Reagent volume
1.0 ml
Wavelength
505 nm (505 to 530 nm, or a green filter)
Flowcell temperature
30 °C or 37 °C
Incubation
10 minutes at 37 °C, or 30 minutes at room temperature (25 ± 5 °C)
Standard
200 mg/dl, included
Reagents
Single liquid reagent, ready to use; lipid clearing factor included
Caution
Contains sodium azide; do not ingest
Storage
2 to 8 °C, until the expiry date on the label
Stock codes
CC3-CLE.05M, CC3-CLE.5MU
Made in
Vadodara, Gujarat, India
Performance2 figures
Linearity
Up to 1000 mg/dl; dilute higher samples with 0.9% saline
Interference
None from hemoglobin up to 375 mg/dl, free bilirubin up to 7.5 mg/dl, lipemia (as triglycerides) up to 1000 mg/dl
Procedure at a glance4 steps
  1. Pipette 1.0 ml of reagent into the blank, standard and test tubes; add 0.01 ml of standard and of sample to their tubes.
  2. Mix and incubate for 10 minutes at 37 °C, or 30 minutes at room temperature (25 ± 5 °C).
  3. Read the standard and the test against the reagent blank at 505 nm.
  4. Cholesterol (mg/dl) = absorbance of test ÷ absorbance of standard × 200.

Every stock code.

Cholesterol-L (Single Liquid) ships in 2 pack sizes, each with its own stock code.

Stock codes and pack sizes for Cholesterol-L (Single Liquid)
Stock code Pack Photograph Datasheet
CC3-CLE.05M 2 x 60 ml Cholesterol-L (Single Liquid), 2 x 60 ml Datasheet PDF, CC3-CLE.05M
CC3-CLE.5MU 5 x 60 ml Cholesterol-L (Single Liquid), 5 x 60 ml Datasheet PDF, CC3-CLE.5MU

Questions and answers.

Cholesterol-L (Single Liquid): what sample does it need?

The datasheet gives the sample as Serum: fresh, clear, fasting, no hemolysis; heparin or EDTA plasma may also be used. The sample volume is 10 µl.

Cholesterol-L (Single Liquid): what method does it use?

The method on the datasheet is Enzymatic CHOD-PAP, end point.

Cholesterol-L (Single Liquid): what wavelength is it read at?

According to the datasheet, it is read at 505 nm (505 to 530 nm, or a green filter).

Cholesterol-L (Single Liquid): what is the linear range?

The datasheet gives the linear range as Up to 1000 mg/dl; dilute higher samples with 0.9% saline.

Cholesterol-L (Single Liquid): how should it be stored?

The datasheet gives storage as 2 to 8 °C, until the expiry date on the label.

Cholesterol-L (Single Liquid): what pack sizes are available?

It comes in 2 pack sizes, each with its own stock code: 2 x 60 ml (CC3-CLE.05M), 5 x 60 ml (CC3-CLE.5MU).

More clinical chemistry.

All 50 clinical chemistry reagents →