Drawing, not a photograph. This pack has not been photographed yet; the stock codes and pack sizes below are the specification.
3 pack sizes, 40 to 240 ml.
Stock code CC3-CEN.024
- Made and batch-released in Vadodara
- ISO 13485:2016 quality system
- Datasheet published: see every stock code below
Product details.
Creatinine (Enzymatic) is a reagent for the in vitro quantitative detection of creatinine in human serum on automated clinical chemistry analyzers, developed to avoid the chromogen interference of the Jaffe method. Creatininase, creatinase and sarcosine oxidase release hydrogen peroxide, which with peroxidase forms a quinoneimine dye read at 546 nm in proportion to creatinine. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.
The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.
- Method
- Enzymatic colorimetric
- Sample
- Fresh serum without hemolysis
- Sample volume
- 3 µl
- Wavelength
- 546 nm (700 nm secondary)
- Linearity
- Up to 17 mg/dl
- Storage
- 2 to 8 °C
Full specification17 items
- Category
- Clinical chemistry
- Pack
- 40 to 240 ml
- Pack sizes
- 3 to choose from
- Datasheet
- Published
- Method
- Enzymatic colorimetric, end point
- Sample
- Fresh serum without hemolysis
- Sample volume
- 3 µl
- Reagent volume
- 150 µl of R1, then 50 µl of R2
- Wavelength
- 546 nm (700 nm secondary)
- Temperature
- 37 °C
- Incubation
- 5 minutes with R1, then 5 minutes after R2
- Calibration
- Two-point; liquid calibrator included
- Reagents
- R1 and R2, liquid, ready to use
- On-board stability
- 30 days
- Storage
- 2 to 8 °C; do not freeze
- Stock codes
- CC3-CEN.024, CC3-CEN.24U, CC3-CEN.24V
- Made in
- Vadodara, Gujarat, India
Performance3 figures
- Linearity
- Up to 17 mg/dl
- Precision
- CV ≤ 5%; inter-batch deviation R ≤ 10%
- Interference
- None up to ascorbic acid 0.5 g/L, bilirubin 0.4 g/L, hemoglobin 5 g/L, triglyceride 10 g/L
Procedure at a glance4 steps
- Mix 3 µl of sample with 150 µl of R1 and incubate for 5 minutes at 37 °C; read absorbance A1.
- Add 50 µl of R2, mix, and incubate for another 5 minutes; read absorbance A2.
- Work out ΔA = A2 − A1 at 546 nm (700 nm secondary).
- Creatinine (mg/dl) = ΔA of sample ÷ ΔA of calibrator × calibrator concentration.
Every stock code.
Creatinine (Enzymatic) ships in 3 pack sizes, each with its own stock code.
| Stock code | Pack | Photograph | Datasheet |
|---|---|---|---|
| CC3-CEN.024 | 2 x 20 ml | — | Datasheet |
| CC3-CEN.24U | 2 x 40 ml | — | Datasheet |
| CC3-CEN.24V | 3 x 80 ml | — | Datasheet |