Photographed: the 12 Test pack, stock code CC1-GPD.09M.
Supplied as 12 tests.
Stock code CC1-GPD.09M
- Made and batch-released in Vadodara
- ISO 13485:2016 quality system
- Datasheet published: see every stock code below
Product details.
G-6-PD (Qualitative) is a visual dye-decolourisation test for screening glucose-6-phosphate dehydrogenase deficiency in erythrocytes. G6PD in a red-cell hemolysate generates NADPH, which with PMS decolourises blue DCPIP dye; the time taken to decolourise is inversely proportional to G6PD activity. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.
The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.
- Method
- Visual dye decolourisation (DCPIP)
- Sample
- Whole blood with EDTA
- Sample volume
- 0.02 ml of whole blood
- Storage
- 2 to 8 °C
- Pack
- 12 tests
- Datasheet
- Published
Full specification15 items
- Category
- Clinical chemistry
- Pack
- 12 tests
- Datasheet
- Published
- Method
- Visual dye decolourisation (DCPIP)
- Sample
- Whole blood with EDTA, not heparin; finger-prick blood if hemoglobin is close to 15 g/dl
- Sample volume
- 0.02 ml of whole blood; more if hemoglobin is below 13.6 g/dl
- Incubation
- 37 °C, sealed under inert oil
- Read at
- 30 minutes, then every 5 minutes until decolourised
- Controls
- Run a known healthy, non-deficient control on opening a kit
- Limitation
- A high reticulocyte count can give a false normal result; vitamin C may interfere
- Reagents
- Co-enzyme substrate, buffer, lysing reagent and inert oil; reconstitute just before use
- Caution
- Use the lysing reagent cold (4 °C); do not disturb the vial after adding the oil
- Storage
- 2 to 8 °C, until the expiry date on the label
- Stock codes
- CC1-GPD.09M
- Made in
- Vadodara, Gujarat, India
Procedure at a glance5 steps
- Add 0.5 ml of buffer to the co-enzyme substrate vial, shake to dissolve, and use within 10 minutes.
- Mix 0.02 ml of whole blood (adjusted for hemoglobin) with 1.0 ml of pre-cooled lysing reagent and keep at 2 to 4 °C for 10 to 15 minutes.
- Transfer the hemolysate into the working reagent, shake, overlay with 1.0 ml of inert oil, seal air-tight, and incubate at 37 °C.
- Watch for the change from blue to brownish colour from 30 minutes, every 5 minutes.
- Normal samples decolourise in 30 to 60 minutes; G6PD-deficient samples take 2 to 24 hours.
Every stock code.
G-6-PD (Qualitative) ships in one pack size.
| Stock code | Pack | Photograph | Datasheet |
|---|---|---|---|
| CC1-GPD.09M | 12 Test | ![]() |
Datasheet |
More clinical chemistry.