CLINICAL CHEMISTRY · G-6-PD (QUANTITATIVE)

G-6-PD (Quantitative)

G-6-PD (Quantitative), 12 x 1.1 ml

Photographed: the 12 x 1.1 ml pack, stock code CC1-GPD.009.

Supplied as 13.2 ml.

Pack size

Stock code CC1-GPD.009

ASK ABOUT THIS PRODUCT →
  • Made and batch-released in Vadodara
  • ISO 13485:2016 quality system
  • Datasheet published: see every stock code below

Product details.

G-6-PD (Quantitative) is a reagent kit for the quantitative kinetic estimation of glucose-6-phosphate dehydrogenase in erythrocytes. The enzyme, released from lysed red cells, oxidises glucose-6-phosphate while reducing NADP to NADPH, and the rise in absorbance at 340 nm is proportional to G6PD activity; an inhibitor prevents interference from 6-phosphogluconate dehydrogenase. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.

The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.

Method
Kinetic UV, at 340 nm
Sample
Whole blood
Sample volume
25 µl (0.025 ml) of hemolysate
Wavelength
340 nm
Storage
2 to 8 °C
Pack
13.2 ml
Full specification17 items
Pack
13.2 ml
Datasheet
Published
Method
Kinetic UV, at 340 nm
Sample
Whole blood: fresh, with citrate, oxalate or heparin; red cells washed and lysed
Sample volume
25 µl (0.025 ml) of hemolysate
Reagent volume
1.0 ml of working reagent
Wavelength
340 nm
Temperature
30 °C
Readings
4, at 60-second intervals after a 180-second delay
Also needed
Hemoglobin and red cell count of the whole blood
Reagents
Co-enzyme substrate vials, buffer and lysing reagent; use the lysing reagent cold (0 to 8 °C)
Working reagent
7 days at 2 to 8 °C
Caution
Contains sodium azide; do not ingest
Storage
2 to 8 °C, until the expiry date on the label
Stock codes
CC1-GPD.009
Made in
Vadodara, Gujarat, India
Procedure at a glance5 steps
  1. Wash 0.1 ml of whole blood 3 times in 2 ml of 0.9% saline; lyse the packed cells in 0.5 ml of cold lysing reagent, keep at 2 to 4 °C for 15 minutes to 2 hours, and centrifuge at 3000 rpm for 5 minutes.
  2. Add 1.1 ml of buffer to a substrate vial, mix, and wait 5 minutes.
  3. Add 0.025 ml of hemolysate to 1.0 ml of working reagent, mix and aspirate.
  4. After 180 seconds, read the absorbance at 340 nm every minute for 3 minutes and work out ΔA/min.
  5. G6PD (U/10¹² RBC) = ΔA/min × 36013 ÷ red cell count (millions/mm³); G6PD (U/g Hb) = ΔA/min × 3601 ÷ Hb (g/dl).

Every stock code.

G-6-PD (Quantitative) ships in one pack size.

Stock codes and pack sizes for G-6-PD (Quantitative)
Stock code Pack Photograph Datasheet
CC1-GPD.009 12 x 1.1 ml G-6-PD (Quantitative), 12 x 1.1 ml Datasheet

More clinical chemistry.