SEROLOGY · WIDAL TUBE TEST (O, H, AH, BH ANTIGEN)

Widal Tube Test (O, H, AH, BH antigen)

Widal Tube Test (O, H, AH, BH antigen), 4 x 50 ml

Photographed: the 4 x 50 ml pack, stock code IS-SAW.78P.

Supplied as 200 ml.

Pack size

Stock code IS-SAW.78P

ASK ABOUT THIS PRODUCT →
  • Made and batch-released in Vadodara
  • ISO 13485:2016 quality system
  • Datasheet published: see every stock code below

Product details.

The Widal Tube Test (O, H, AH, BH antigen) is a tube agglutination kit for detecting and titrating antibodies to Salmonella in serum, for the diagnosis of enteric fever. Killed suspensions of S. typhi O and H and S. paratyphi A (H) and B (H) are mixed with a serial dilution of the patient's serum and incubated overnight. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.

The specifications and steps on this page are from the datasheet, last updated 04-2023. The datasheet governs: read it before use, or ask us.

Method
Tube agglutination
Sample
Fresh serum preferred
Storage
2 to 8 °C
Pack
200 ml
Full specification13 items
Category
Serology
Pack
200 ml
Datasheet
Published
Method
Tube agglutination, killed Salmonella antigens
Antigens
S. typhi O and H, S. paratyphi A (H) and B (H), 50 ml each
Sample
Fresh serum preferred; store at 2 to 8 °C away from light and test within 24 hours
Dilutions
1:30, 1:60, 1:120, 1:240 and 1:480, with a saline control
Incubation
Overnight (16 to 20 hours) at 37 °C
Reading
'H' agglutination is floccular, like cotton wool; 'O' is fine, granular and matted at the bottom of the tube
Limitation
Non-specific: confirm doubtful results by microbiology; past infection or TAB vaccination can raise titres, and a rising titre on repeat testing points to true infection
Storage
2 to 8 °C, away from direct light
Stock codes
IS-SAW.78P
Made in
Vadodara, Gujarat, India
Procedure at a glance5 steps
  1. For each serum, set 4 rows of 6 tubes (3 × 3/8 inch) in a rack.
  2. Make a master dilution in 5 tubes: 7 ml of normal saline in the first and 3.5 ml in the rest; add 0.5 ml of serum to the first and carry 3.5 ml down the row, for 1:30 to 1:480.
  3. Transfer 0.5 ml from each master tube to the matching tube in every row, and put 0.5 ml of saline in the sixth tube of each row as a control.
  4. Add 0.5 ml of S. typhi O, S. typhi H, S. paratyphi A (H) and B (H) antigen to rows 1 to 4, shake, and incubate at 37 °C overnight (16 to 20 hours).
  5. The titre is the highest dilution with visible agglutination, judged against the saline control.

Every stock code.

Widal Tube Test (O, H, AH, BH antigen) ships in one pack size.

Stock codes and pack sizes for Widal Tube Test (O, H, AH, BH antigen)
Stock code Pack Photograph Datasheet
IS-SAW.78P 4 x 50 ml Widal Tube Test (O, H, AH, BH antigen), 4 x 50 ml Datasheet

More serology.