RAPID DIAGNOSTICS · DENGUE ANTIGEN NS1 CARD

Dengue Antigen NS1 Card

Cassette: no photograph of this pack is published yet

Drawing, not a photograph. This pack has not been photographed yet; the stock codes and pack sizes below are the specification.

CARD format. 2 pack sizes, 10 to 25 tests.

Choose a pack size

Stock code RDT-DEG.101M

ASK ABOUT THIS PRODUCT →
  • Made and batch-released in Vadodara
  • ISO 13485:2016 quality system
  • Datasheet published: see every stock code below

Product details.

The Dengue Antigen NS1 Card is a rapid, qualitative solid-phase immunochromatographic test for dengue virus NS1 antigen in human serum or plasma. Anti-dengue NS1 monoclonal antibody on the test line captures the NS1-gold complex; NS1 is found from the first day and up to 9 days after the onset of fever. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.

The specifications and steps on this page are from the datasheet, last updated 04-2026. The datasheet governs: read it before use, or ask us.

Method
Chromatographic immunoassay
Detects
Dengue virus NS1 antigen
Sample
Serum or plasma
Sample volume
2 to 3 drops (50 to 75 µl)
Read at
15 to 20 minutes
Storage
2 to 40 °C
Full specification14 items
Format
CARD
Pack
10 to 25 tests
Pack sizes
2 to choose from
Datasheet
Published
Method
Chromatographic immunoassay, colloidal gold
Detects
Dengue virus NS1 antigen
Sample
Serum or plasma (heparin, EDTA or citrate); 3 days at 2 to 8 °C, or frozen at −20 °C beyond that
Sample volume
2 to 3 drops (50 to 75 µl)
Read at
15 to 20 minutes; not after 20 minutes
Limitation
A primary screening test; a negative result does not rule out infection
Storage
2 to 40 °C in the sealed pouch
Stock codes
RDT-DEG.101M, RDT-DEG.101MU
Made in
Vadodara, Gujarat, India
Procedure at a glance5 steps
  1. Bring the specimen and the device to room temperature (15 to 30 minutes).
  2. Open the pouch and use the device straight away.
  3. Add 2 to 3 drops (50 to 75 µl) of serum or plasma to sample well S.
  4. Read at 15 to 20 minutes, not later.
  5. C only is negative; C and T is positive for dengue NS1 antigen; no C line, or T only, means the test is invalid.

Every stock code.

Dengue Antigen NS1 Card ships in 2 pack sizes, each with its own stock code.

Stock codes and pack sizes for Dengue Antigen NS1 Card
Stock code Pack Photograph Datasheet
RDT-DEG.101M 10 Test Datasheet
RDT-DEG.101MU 25 Test Datasheet