HCV Card, 50 Test HCV Card, 50 Test, view 2

Photographed: the 50 Test pack, stock code RDT-HCC.105.

Card format. Supplied as 50 tests.

Pack size

Stock code RDT-HCC.105

  • Made and batch-released in Vadodara
  • ISO 13485:2016 quality system
  • Datasheet published: see every stock code below

Product details.

Intended use: The HCV Card is a rapid chromatographic immunoassay for the qualitative detection of antibodies to hepatitis C virus in human serum or plasma. It uses the double antigen sandwich principle, with recombinant HCV core, NS3, NS4 and NS5 antigens on the test line and a goat anti-mouse IgG control line. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.

The specifications and steps on this page are from the datasheet, last updated 01-2026. The datasheet governs: read it before use, or ask us.

Method
Double antigen sandwich
Sample
Serum or plasma
Sample volume
1 drop (about 30 µl)
Read at
Within 15 minutes
Storage
2 to 40 °C
Format
Card
Full specification15 items
Format
CARD
Pack
50 tests
Datasheet
Published
Method
Double antigen sandwich, lateral flow
Detects
Antibodies to HCV (core, NS3, NS4 and NS5 antigens)
Sample
Serum or plasma, tested soon after separation; 3 days at 2 to 8 °C, or frozen at −20 °C beyond that
Sample volume
1 drop (about 30 µl)
Buffer
2 drops (about 60 µl), added straight after the sample
Read at
Within 15 minutes; strong positives may show by 5 minutes; not after 20 minutes
Limitation
Shows only the presence or absence of HCV antibody; if negative with persisting symptoms, repeat on a fresh sample
Storage
2 to 40 °C in the sealed pouch
Caution
Handle specimens as infectious; soak used material in 0.5 to 1% sodium hypochlorite for 1 hour
Stock codes
RDT-HCC.105
Made in
Vadodara, Gujarat, India
Procedure at a glance4 steps
  1. Bring the card and the sample to room temperature, and lay the card flat.
  2. Add 1 drop (about 30 µl) of serum or plasma, then 2 drops (about 60 µl) of buffer, to the sample well without overflowing.
  3. Read within 15 minutes; do not read after 20 minutes.
  4. C only is negative; C and T is positive for HCV antibody, whatever the line intensities; no C line means the test is invalid.

Every stock code.

HCV Card ships in one pack size.

Stock codes and pack sizes for HCV Card
Stock code Pack Photograph Datasheet
RDT-HCC.105 50 Test HCV Card, 50 Test Datasheet PDF, RDT-HCC.105

Questions and answers.

HCV Card: what sample does it need?

The datasheet gives the sample as Serum or plasma, tested soon after separation; 3 days at 2 to 8 °C, or frozen at −20 °C beyond that. The sample volume is 1 drop (about 30 µl).

HCV Card: what method does it use?

The method on the datasheet is Double antigen sandwich, lateral flow.

HCV Card: how should it be stored?

The datasheet gives storage as 2 to 40 °C in the sealed pouch.

HCV Card: what pack sizes are available?

It comes in one pack size: 50 Test (RDT-HCC.105).