Skip to content
ENQUIRY BASKET · 0
Empty. Add products with the + on any card, then send the whole list as one enquiry.
IMMUNOTURBIDIMETRIC · RF TURBILATEX

RF Turbilatex

RF Turbilatex, 1 x 50 ml

Photographed: the 1 x 50 ml pack, stock code IM-RFT.083.

Supplied as 50 ml.

Pack size

Stock code IM-RFT.083

ASK ABOUT THIS PRODUCT →
  • Made and batch-released in Vadodara
  • ISO 13485:2016 quality system
  • Datasheet published: see every stock code below

Product details.

RF Turbilatex is a quantitative latex turbidimetric test for rheumatoid factors (RF) in human serum or plasma. Latex particles coated with human gammaglobulin agglutinate with the RF in the sample, and the change in absorbance is read against a multi-point or single-point calibration. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.

The specifications and steps on this page are from the datasheet, last updated 06-2023. The datasheet governs: read it before use, or ask us.

Method
Latex turbidimetry
Sample
Fresh serum
Sample volume
7 µl
Wavelength
650 nm (600 to 650 nm)
Linearity
Up to 100 IU/ml single point
Measuring range
20 to 150 IU/ml on the calibration curve
Full specification17 items
Pack
50 ml
Datasheet
Published
Method
Latex turbidimetry, end point (multi-standard)
Sample
Fresh serum: stable 7 days at 2 to 8 °C or 3 months at −20 °C; centrifuge samples with fibrin; not highly hemolysed or lipemic
Sample volume
7 µl
Reagent volume
1 ml of working reagent (0.9 ml of R1 + 0.1 ml of R2)
Wavelength
650 nm (600 to 650 nm)
Temperature
37 °C
Incubation
2 minutes
Calibration
6-point curve (20 to 150 IU/ml), or single point with the calibrator diluted to 45 IU/ml
Working reagent
30 days at 2 to 8 °C
Calibrator stability
1 month at 2 to 8 °C; 3 months at −20 °C
Caution
Human-origin components tested negative for HBsAg, HCV and HIV; handle as potentially infectious
Storage
2 to 8 °C; do not freeze
Stock codes
IM-RFT.083
Made in
Vadodara, Gujarat, India
Performance5 figures
Linearity
Up to 100 IU/ml single point; up to 150 IU/ml multi-point
Measuring range
20 to 150 IU/ml on the calibration curve; dilute higher samples 1/5 in 9 g/L NaCl
Detection limit
3 IU/ml
Sensitivity
Δ 3.34 mA per IU/ml
Interference
None from bilirubin 20 mg/dl, hemoglobin 10 g/L, lipemia 10 g/L
Procedure at a glance5 steps
  1. Mix 0.9 ml of diluent (R1) with 0.1 ml of latex (R2) as the working reagent.
  2. Prepare the 6 calibrator dilutions in 9 g/L NaCl for a multi-point curve, or mix 30 µl of calibrator with 70 µl of NaCl for a single point (45 IU/ml).
  3. Add 7 µl of calibrator or sample to 1.0 ml of working reagent (blank: 7 µl of NaCl) and mix.
  4. Read the blank (A1) and the sample or calibrator (A2) at 650 nm after 2 minutes.
  5. Read RF off the calibration curve of (A2 − A1), or RF (IU/ml) = (A2 − A1) of sample ÷ (A2 − A1) of calibrator × diluted calibrator concentration.

Every stock code.

RF Turbilatex ships in one pack size.

Stock codes and pack sizes for RF Turbilatex
Stock code Pack Photograph Datasheet
IM-RFT.083 1 x 50 ml RF Turbilatex, 1 x 50 ml Datasheet

More immunoturbidimetric.

All 7 in immunoturbidimetric →

Tell us what
your lab needs.

ENQUIRE NOW →