Drawing, not a photograph. This pack has not been photographed yet; the stock codes and pack sizes below are the specification.
Supplied as 60 ml.
Stock code IM-FER.079
- Made and batch-released in Vadodara
- ISO 13485:2016 quality system
- Datasheet published: see every stock code below
Product details.
Ferritin Turbilatex is a reagent kit for the in vitro quantitative detection of ferritin in human serum on semi-automated and fully automated clinical chemistry analyzers. Ferritin and latex-bound antibody form a complex whose turbidity is proportional to the ferritin present. Manufactured at Reckon’s Vadodara facility under ISO 13485:2016 quality management, and released against reference standards before dispatch.
The specifications and steps on this page are from the datasheet, last updated 08-2023. The datasheet governs: read it before use, or ask us.
- Method
- Latex turbidimetry
- Sample
- Fresh serum without hemolysis
- Sample volume
- 20 µl semi-automated
- Wavelength
- 570 nm
- Linearity
- 10 to 750 ng/ml (R ≥ 0.990)
- Storage
- 2 to 8 °C
Full specification15 items
- Category
- Immunoturbidimetric
- Pack
- 60 ml
- Datasheet
- Published
- Method
- Latex turbidimetry, end point (increase)
- Sample
- Fresh serum without hemolysis
- Sample volume
- 20 µl semi-automated; 7 µl fully automated
- Reagents
- R1 diluent 2 × 20 ml, R2 latex 2 × 10 ml, calibrator 4 × 0.5 ml; liquid, ready to use
- Wavelength
- 570 nm
- Temperature
- 37 °C
- Incubation
- 5 minutes, then 5 minutes after R2
- Calibration
- Multi-point, calibrator included
- On-board stability
- 30 days
- Storage
- 2 to 8 °C; do not freeze
- Stock codes
- IM-FER.079
- Made in
- Vadodara, Gujarat, India
Performance3 figures
- Linearity
- 10 to 750 ng/ml (R ≥ 0.990)
- Precision
- CV below 10%
- Interference
- None significant up to bilirubin 85 mg/dl, hemoglobin 300 mg/dl, triglyceride 1000 mg/dl
Procedure at a glance5 steps
- Pipette 400 µl of R1 into the blank, calibrator and test tubes.
- Add 20 µl of calibrator to the calibrator tube and 20 µl of sample to the test tube.
- Mix and incubate for 5 minutes at 37 °C.
- Add 200 µl of R2 to every tube, mix, and incubate for 5 minutes at 37 °C.
- Read the test and calibrator against the reagent blank at 570 nm, and read ferritin off the calibration curve.
Every stock code.
Ferritin Turbilatex ships in one pack size.
| Stock code | Pack | Photograph | Datasheet |
|---|---|---|---|
| IM-FER.079 | 2 x 30 ml | — | Datasheet |